Iatrogenic uterine rupture caused by overinflation of RUMI manipulator balloon
作者: Hsin-hung WuGuang-Perng YehTsung-Che Hsieh
作者单位: 1Department of Obstetrics and Gynecology, Changhua Christian Hospital, Changhua
2College of Health Sciences, Institute of Medical Research, Chang Jung Christian University, Tainan
3Department of Obstetrics and Gynecology, Chung Shan Medical University, Taichung, Taiwan.
刊名: The Journal of Minimally Invasive Gynecology, 2005, Vol.12 (2), pp.174-176
来源数据库: Elsevier Journal
DOI: 10.1016/j.jmig.2005.01.012
关键词: RUMIUterine ruptureOverinflation
原始语种摘要: The RUMI uterine manipulator is a useful device for uterine mobilization and chromopertubation during laparoscopic surgery. We report two cases of iatrogenic uterine rupture caused by overinflation of the intrauterine balloon of the RUMI manipulator during chromopertubation. One rupture, which occurred on the uterine fundus, simply was sutured during surgery. The other rupture, which occurred into the uterine lateral wall, caused retroperitoneal hematoma. A second operation was required to treat this complication. Overinflation of the tip balloon of the RUMI manipulator during chromopertubation can cause uterine rupture that can result in massive hematoma.