Comparison between new-generation SiPM-based and conventional PMT-based TOF-PET/CT
作者: Kei WagatsumaKenta MiwaMuneyuki SakataKeiichi OdaHaruka OnoMasashi KameyamaJun ToyoharaKenji Ishii
作者单位: 1Research Team for Neuroimaging, Tokyo Metropolitan Institute of Gerontology, Tokyo, Japan
2School of Health Science, International University of Health and Welfare, Ohtawara, Japan
3Faculty of Health Science, Hokkaido University of Science, Sapporo, Japan
4Department of Diagnostic Radiology, Tokyo Metropolitan Geriatric Hospital, Tokyo, Japan
刊名: Physica Medica, 2017, Vol.42 , pp.203-210
来源数据库: Elsevier Journal
DOI: 10.1016/j.ejmp.2017.09.124
关键词: Silicon-photomultiplierNational Electrical Manufactured AssociationInstrumentationPhysical performanceImage quality
原始语种摘要: Abstract(#br)Purpose(#br)This study aimed to determine whether the SiPM-PET/CT, Discovery MI (DMI) performs better than the PMT-PET/CT system, Discovery 710 (D710).(#br)Methods(#br)The physical performance of both systems was evaluated using NEMA NU 2 standards. Contrast (%), uniformity and image noise (%) are criteria proposed by the Japanese Society of Nuclear Medicine (JSNM) for phantom tests and were determined in images acquired from Hoffman and uniform phantoms using the DMI and D710. Brain and whole-body [ 18 F]FDG images were also acquired from a healthy male using the DMI and D710.(#br)Results(#br)The spatial resolution at 1.0cm off-center in the DMI and D710 was 3.91 and 4.52mm, respectively. The sensitivity of the DMI and D710 was 12.62 and...
全文获取路径: Elsevier  (合作)