Identification of a matrix effect in the MC-ICP-MS due to sample purification using ion exchange resin: An isotopic case study of molybdenum
作者: Aaron J. PietruszkaAudrey D. Reznik
作者单位: 1Department of Geological Sciences, San Diego State University, San Diego, CA 92182, USA
刊名: International Journal of Mass Spectrometry, 2007, Vol.270 (1), pp.23-30
来源数据库: Elsevier Journal
DOI: 10.1016/j.ijms.2007.11.001
关键词: MolybdenumIsotopeMass spectrometryMC-ICP-MSMatrix effect
原始语种摘要: Abstract(#br)Multiple-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) has become the preferred method for precise and accurate measurements of the relative abundances of many radiogenic and stable isotopes in natural materials. Isotopic analyses by MC-ICP-MS require a correction for instrumental mass-dependent isotopic fractionation (“instrumental mass bias”). Two techniques have been used to correct for instrumental mass bias in the MC-ICP-MS: (1) standard-sample bracketing (SSB) or (2) double spiking. SSB is often cited as the preferred method, but it is more susceptible to matrix effects. Here we demonstrate that a matrix effect in the MC-ICP-MS may arise indirectly from the chemical separation and purification of molybdenum using anion exchange resin. The results of...