THE EFFECT OF ANTICOAGULANT TYPES ON THE IN VITRO ANALYSIS OF CLOPIDOGREL IN HUMAN PLASMA USING LIQUID CHROMATOGRAPHY TANDEM-MASS SPECTROMETRY
作者: Yahdiana HarahapAnisa MaulidinaDelly Ramadon
作者单位: 1Faculty of Pharmacy, Universitas Indonesia, Depok 16424, Indonesia. yahdiana03@yahoo.com
刊名: International Journal of Applied Pharmaceutics, 2018, Vol.10 (1), pp.392-395
来源数据库: Innovare Academics Sciences PVT.Ltd.
DOI: 10.22159/ijap.2018.v10s1.87
原始语种摘要: Objective: The aim of this study was to optimize and validate a plasma clopidogrel analysis method using liquid chromatography tandem-massspectrometry.Methods: Plasma samples were analyzed using a BEH C18 column (1.7 μm; 100 mm×2.1 mm), the mobile phase was 0.1% formic acid in acetonitrile(30:70, v/v). The flow rate was 0.2 mL/min, with a column temperature set to 35°C, an injection volume of 5 μL, an analysis time of 4 min, andirbesartan as the internal standard. Aliquots were obtained by liquid-liquid extraction using ammonium acetate and diethyl ether. The stability andpeak area ratio of the respective plasma area responses were evaluated using ANOVA.Results: No significant differences (p>0.05) were observed between anticoagulants regarding analyte stability. However, the peak area...