Biomechanical Characterization of Failure the 316L Stainless Steel for Femoral Compression Plates
作者: Abdelyamine BoukhobzaMourad BriouaSmail BenaichaKamel Fedaoui
作者单位: 1University of Batna
刊名: Journal of Biomimetics, Biomaterials and Biomedical Engineering, 2017, Vol.4684 , pp.68-74
来源数据库: Trans Tech Journal
DOI: 10.4028/www.scientific.net/JBBBE.34.68
关键词: 316L Austenitic Stainless SteelBone Fixture Compression PlateBreakage Cause AnalysisBreakage Surface Analysis
原始语种摘要: The aim of this paper is to examine a premature breakage of two compression plates for fixing broken bones with different patients for the period of their recovery. Each compression plate's breakage can induce grave consequences such as a new surgery, unexpected undesired complications and a prolonged healing time. The investigation of the compression plate breakage causes required an examination of the chemical composition and steel hardness, metallographic examination as well as that of the compression plate breakage surface by means of macroscopic and microscopic observations using microscope. On the origin of the results it can be established that the breakage was caused by high static load.
全文获取路径: PDF下载  Trans Tech 
分享到:

×
关键词翻译
关键词翻译
  • compression 压缩
  • breakage 崩落
  • undesired 不希望有的
  • examination 审查
  • Analysis 分析
  • hardness 硬度
  • investigation 研究
  • metallographic 金相学的
  • static 静止的
  • unexpected 突然的