FLUCTUATION OF INFUSION PRESSURE DURING MICROINCISION VITRECTOMY USING THE CONSTELLATION VISION SYSTEM
作者: Yong Joon KimSun Ho ParkKyung Seek Choi
作者单位: 1Department of Ophthalmology, Soonchunhyang University College of Medicine, Soonchunhyang University Seoul Hospital, Seoul, Republic of Korea.
刊名: Retina, 2015, Vol.35 (12), pp.2529-2536
来源数据库: Wolters Kluwer Journal
DOI: 10.1097/IAE.0000000000000625
原始语种摘要: PURPOSE:: To measure fluctuations in infusion pressure and intraocular pressure (IOP) during vitrectomy performed using a flow-based IOP control system. METHODS:: Using 3 vitrectomized porcine eyes, the authors simultaneously measured infusion pressure and IOP during vitreous cutting and aspiration and after extraction of operative instruments in 23-gauge and 25-gauge system. The measurements were performed with the “IOP control” setting turned on or off. The efficacy of valved cannula and a built-in “IOP control limit” module in attenuation of infusion pressure fluctuation was evaluated. RESULTS:: At set pressure of 30 mmHg and 60 mmHg, the mean infusion pressure levels were 43.7 mmHg and 78.7 mmHg in the vitreous cutting mode, 67.4 mmHg and 101.2 mmHg in the aspiration mode, and 72.8...