Effects of a novel tungsten-impregnated rubber neck shield on the quality of cerebral images acquired using 15 O-labeled gas
作者: Kei WagatsumaKeiichi OdaKenta MiwaMotoki InajiMuneyuki SakataJun ToyoharaKiichi IshiwataMasayuki SasakiKenji Ishii
作者单位: 1Research Team for Neuroimaging, Tokyo Metropolitan Institute of Gerontology
2Kyushu University
3Hokkaido University of Science
4International University of Health and Welfare
5Tokyo Medical and Dental University
6Institute of Cyclotron and Drug Discovery Research, Southern TOHOKU Research Institute for Neuroscience
7Fukushima Medical University
刊名: Radiological Physics and Technology, 2017, Vol.10 (4), pp.422-430
来源数据库: Springer Nature Journal
DOI: 10.1007/s12194-017-0414-5
关键词: Positron emission tomographyOxygen-15Image qualityCount characteristicsNeck shield
英文摘要: The present study aimed to validate the effects of a novel tungsten-impregnated rubber neck shield on the quality of phantom and clinical 15O-labeled gas positron emission tomography (PET) images. Images were acquired in the presence or absence of a neck shield from a cylindrical phantom containing [15O]H2O (phantom study) and from three individuals using [15O]CO2, [15O]O2 and [15O]CO gas (clinical study). Data were acquired in three-dimensional (3D) mode using a Discovery PET/CT 710. Values for cerebral blood flow, cerebral blood volume, oxygen extraction fraction, and cerebral metabolic rate of oxygen with and without the neck shield were calculated from 15O-labeled gas images. Arterial...
原始语种摘要: The present study aimed to validate the effects of a novel tungsten-impregnated rubber neck shield on the quality of phantom and clinical 15O-labeled gas positron emission tomography (PET) images. Images were acquired in the presence or absence of a neck shield from a cylindrical phantom containing [15O]H2O (phantom study) and from three individuals using [15O]CO2, [15O]O2 and [15O]CO gas (clinical study). Data were acquired in three-dimensional (3D) mode using a Discovery PET/CT 710. Values for cerebral blood flow, cerebral blood volume, oxygen extraction fraction, and cerebral metabolic rate of oxygen with and without the neck shield were calculated from 15O-labeled gas images. Arterial...
全文获取路径: Springer Nature  (合作)